The issue seems resolved. Loosening the occlusion microswitch and adjusting to increase the tension slightly on the detect finger. a couple grams would activate and stay activated before. Of course, a mechanical device with an activated limit switch (occlusion) would not start a rotation test... ICU Medical LifeCare PCA Mechanical Overview